CCD2 Information


CCD Name

CCD2

Serial Number

 

Description

Loral 64x64 Frame Transfer CCD, front-illuminated,

Pixel Size(micron)

15x15

Chip size

128x64 (0 prescan pixels)

Preflash, e-

None

Readout Mode

 

NON-MPP

Dark Current (-40 DegC)

 

3e-/pixel/sec

Full Well Capacity [e­/px]

 

10K


Astromed 3200 Controller Performance

 
Amplifier bottom left in use.
 
 
 

Characteristic

Amplifier B

Noise (e-/RMS)

4.5

Gain (e-/ADU)

1.0

Full Frame READOUT TIME [secs]

6.7m

Control Temperature [K]

220

Linearity 0.1% [e­]

 

Dark Current [e­/px/hr]

 


SDSU1 Controller Performance

 
Amplifier bottom left in use.
 

Characteristic

Amplifier B

Noise (e-/RMS)

 

Gain (e-/ADU)

 

Full Frame READOUT TIME [secs]

 

Control Temperature

220K

Linearity 0.1% [e­]

 

Dark Current [e­/px/hr]

 


Amplifier Screening Data

Setup

 
Data Rate       : CCDI 120       VO Load: IDS=0.2mA
Temperature     : 183K 
tc              : 3              pgain  : 16
 
 

DC BIAS

Volts

Clocks

Voltage Swing

VDDA,B

+17.5V

Reset Gate

-8.2 to +1.5

VRDA,B,OFD

+9.0V

Serial Gates

-8.5 to +1.5

VLGA

-6.5V

Summing Well

-8.5 to +1.5

SUB,PKGGND,AMPGND

-4.5V

Parallel 1,2,3

-12.5 to +1.0

 

 

 

 

Amplifier Characteristics

Characteristic

Amplifier A

Amplifier B

Noise (e-/RMS)

 

 

Gain (uV/e-)

 

 

Parallel CTE

 

 

Serial CTE

 

 

 
 
 

Cosmetics

Dead Columns :

Hot Columns :

Low Columns :


Quantum Efficiency

some sort of graph or table


Images

The following images are available as fits files


Detector History

Science grade chip for 2.3m Telescope Tip-tilt

The image problems were fixed by replacing CCD with spare ccd serial number 9. Serial #9 has both bottom amplifiers operational and has a readnoise of 5e at a gain of 1e/ADU and a well depth > 65000. None of the other amplifiers of the original ccd seem to be operational. Since we are still using the bottom left amplifier no code needs to be modified.