Prepared by: John Shobbrook
Whilst in Options → Preferences, select É
q
Show messages
window at startup
q
Reset number
of exposures to 1 after each sequence of >1 exposures
q
Toggle binning
when switching from no-save to save
(The
left monitor is Ô1Õ – the right monitor is Ô0Õ)
á
Focusing
Parameters É Steps 5 First focus position 0.001 Position step 1.000
á
Star profile
radius (pixels) 10
á
Click on Apply
á
At the top of the
Preferences Window, click on General, click on Image Display
and Scaling
Scaling Algorithm
É
Adaptive Median Sample size = 1000 %of image to
readout before scaling = 25
Contrast adjustment = 0.50 Num SDs tolerance before
rescaling = 0.10
WFI
(N.B. Take Dome Flats
before filling the dewar to avoid LN2 spillage).
N.B. If you need to use the shutter windscreen, the ÔraiseÕ and ÔlowerÕ control buttons are located on the dome beneath the shutter opening.
BIAS FRAMES
á
In the Cicada
control window setÉ
á
Object Name: BIAS
á
Exposure
Type: □ BIAS -
□ Save and Display
WFI
á
Exposure Options
- number of exposures: 5
á
Click on Options, click on Show Regions Window
Check that Ôregion 1Õ
is highlighted with a blue box and that the X Y Width and Height settings are
set for the full CCD.
X-offset (XO) = 0,
Y-offset (YO) = 0,
Width = 8392, Height
= 8272.
á
Set the Overscan
Columns to 0 (This Overscan option is for the 2.3m DBS Astromed
controllers. The WFI SDSU
controllers have overscan and bias regions preset).
á
Click on Dismiss - Click
on Expose
Bias
frames should display between 2,500 ~ 4,500 counts
(The
standard deviation of bias frames can be calculated by using GIT -
Load the bias frame into ximtool
– using the left
mouse button drag a box onto the image
- click on the
Statistics
Icon (Bar Graph) - the standard deviation should be
between 2 and 3. Clear the box by
right clicking on the box and choosing ÔdestroyÕ.)
FLAT FIELDS
Sky
Flat preparationÉ
These
should be taken about 15 to 30 minutes after sunset or before sunrise. The key to this is to be about 20¼ away
from the zenith in the direction opposite to where the Sun is setting or
rising. Sky Flats tend to be
better than Dome Flats. As a
general guide É
á
In the late
afternoon, slew the telescope in R.A. away from the sunset to the Sidereal Time
+ 1 hour. In the early morning slew to the
Sidereal Time – 1 hour. In summer slew in Declination to –
20¼. In winter slew in Dec to – 40¼, in spring and autumn slew to about –
30¼
Dome
Flat preparation É
(N.B. Take dome flats before filling the
dewar to avoid LN2
spillage.)
á
In the E.T.S. Configuration
– Configure menu, set
the Dome Control to Manual
á
Enter the dome,
ensure that the console Auto – Slew/Set knob is switched to Slew/Set
á
Using the
Declination – Slew ÔSÕ
button, lower the telescope to about -78¼ (beyond the SCP) (Telescope ZD = 70¡ as shown on the ETS monitor)
á
Rotate the dome
to an Azimuth of 70¡ (as shown on the ETS monitor) or align the red and black
mark on the dome with the marks on the wall behind the dome console.
á
Rotate the dome
so that the dome flat screen and telescope are in alignment.
á
Partially open
the upper shutter to faintly illuminate the screen, turn off all dome lights.
á
If you need to
use dome lights, use the tungsten lights only!
Cicada
settings for Sky/Dome flats É
á
In Options
→ Show Regions Window
use the X Y Width and Height settings for the full CCD. Set x-offset = 0, y-offset = 0,
Width = 8392, Height = 8272.
á
Object Name: Sky_Flat or Dome_Flat,
Exposure Type: □ Flat
á
Start taking
Flats with Exposure Options
set to □ Display Only
á
WFI
á
Set the Overscan
Columns to 0 (This
Overscan option is for the 2.3m DBS Astromed controllers. The WFI SDSU controllers have overscan
and bias regions preset).
á
For Dome Flats
start with an exposure duration of between 10 ~ 25 seconds.
á
Read the counts
in Ximtool and adjust the exposure duration or light entering the dome until
you obtain a value of about 30,000.
á
When the correct
number of counts have been reached, set the Exposure Options to É □
Save and Display
á
Click on Expose – take at least 5 flat fields with each
filter in use.
á
With Sky
Flats, offset the telescope between exposures by about 2 arcmin, by switching
the control room paddle to Set and pressing the ÔNÕ button for about one second.
á
Aim for
brightness counts in the vicinity of 10,000 ~ 30,000.
DARK
FRAMES (Dark current is small and dark
frames are generally not necessary).
POINTING THE TELESCOPE – CALIBRATING THE POINTING
á
Ensure that the Auto
– Slew/Set rotary knob
on the dome console is set to Auto and that the dome floor is fully lowered.
á
Using the E.T.S.
Configuration – Configure
menu, set the Dome Control
to Auto
á
Click on the Motion drop down menu, click on Track
á
Enter the Name
(Optional) R.A., Dec and Epoch (Optional) of
the target object.
á
Click on Track -
Take an exposure to
ensure that the target has been centred.
á
Click on Configuration
– Calibrate Pointing,
enter the R.A. and Dec of the target, click on Calibrate, click on Close
FOCUSING THE TELESCOPE
(Using the WFI Automated Focusing sequence)
á
In Cicada under
the Actions drop down menu,
click on Do_Focus_Sequence
á
Follow the
prompts or refer to p. 17 of the
Cicada UserÕs Manual
á
To highlight a
star, drag across it with the left mouse button then click the right-moost icon
in Cicada ÔRead Marked Region from Image ToolÕ
á
Move the focus
using the ÒInÓ and ÒOutÓ buttons on the round paddle by 0.03 - 0.05 units
between each image, starting at about 31.70 (as displayed on the small red LED
focus readout box)
(Using the MaxIm DL autoguider CCD É)
á
In the Focus
Tab check the Continuous box, then click the Reset button.
á
Ensure that the
Subframe is set to Left = 0 Top
= 0 Width = 320 Height = 240
á
Click the Auto button, this will place a box around the
brightest star in the field.
á
Click on Start
Focus exposures will
continuously cycle. Click on the Inspect tab.
WFI
á
Focus the
telescope using the In and
Out buttons on the control
room hand paddle.
N.B. The ÔInÕ button increases the focus position value, (as displayed on the control room focus position digital display) the ÔOutÕ button decreases the value.
The ÔInspectÕ Tab will
display the information needed to focus the telescope including the FWHM in pixels and the amplitude (ÔMaxÕ
window) of the brightest pixel.
Best focus is obtained with the FWHM value as small as possible and the
ÔMaxÕ value as large as possible.
The narrower the 3-D isometric plot, the better the focus.
The plate scale of
the WFI is 0.375 arcsec per pixel, 25 arcsec per mm or 2.5 pixels per sec of
arc e.g. if a star image is 5
pixels across then the seeing is 2 arcsec. The field of view covers 52 arcminutes with the diagonal
being 1.2 degrees.
You can also monitor focus as you work by
keeping an eye on the images. The
40-inch has astigmatism such that images are elongated at about 45 degrees from
NW to SE when the focus setting is too high and from NE to SW when the focus is
too low.
You can use this astigmatism to focus the
telescope:
If the star images are elongated É
NW-SE => focus is too high: Press the ÒOutÓ button (LED readout
decreases)
NE-SW => focus is too low: Press the ÒInÓ button (LED readout
increases)
Typical focus values are in the range from
31.65 – 31.88
(Using the main WFI CCD
mosaic É)
á
Using the control
room hand paddle, slew the telescope to the South of the chosen focus field
then back North to cancel gear backlash
- stop with the Declination
seconds readout on Ò00Ó (use the
ÔGuideÕ option for fine control).
á
Using the hand
paddle, cancel focus backlash by increasing the focus readout with the IN FOCUS button until you are beyond your start
position. Press the OUT focus button to come back down to your start
position (say 32.40).
á
Set Exposure
Type to É □ Object and
□ Display Only
á
.Set Exposure
Options to É Exposures 1 Duration 10 seconds
□
No pre-flush
and □ No readout
á
Click on Flush to clear the previous image from the CCD array
then click on Expose
á
Using the ÔNÕ paddle button set to Guide move the telescope by 20Ó of arc (watch Dec
change on the E.T.S monitor, declination will decrease as the telescope moves
North).
á
Lower the focus
readout by 0.05 by using the OUT paddle button, record the focus position.
á
Take a second 10
second duration exposure by clicking on Expose
á
Using the ÔNÕ paddle button move the telescope by another
20Ó of arc.
á
Lower the focus
readout by 0.05 by using the OUT paddle button, record focus position.
á
Take another 10
second exposure.
WFI
á
REPEAT the
telescope move, focus and expose steps until you have taken 6 exposures at
differing focus positions. (To
clearly identify between the first and last image in the subsequent readout, you
could also slight shift the last image in R.A.)
á
For the final
exposure CANCEL the Ô
□ NO READOUTÕ option in Exposure
Options
á
Examine the
displayed image, the exposure sequence will be from right to left, (if you used the ÔNÕ button to move the
telescope) note which focus position provides the best image – reset the
focus position accordingly – repeat the focus sequence if necessary. If the image is ÒstreakedÓ re-boot
Cicada.
AUTOGUIDING
á
If the autoguider
PC has to be restarted:
1. When asked for a password, just click ÔCancelÕ
2. The ÔVNC serverÕ is m40slave:0
Password is in the back of the 40Ó CCD Temp Log Book.
3. After Windows starts, click on the ÔMaxIm DLÕ
icon.
4. In ÔMaxIm DLÕ click on Connect to connect to the camera.
5. Scroll down in the MaxIm DL window so that you
can click on ÔGuide CCDÕs – WFIÕ.
á
In the Cicada Tools menu, click on Start GSFind this will open the ÔGuide Star FinderÕ – this shows an outline of the main
and guiding CCDs relative to GSC stars.
á
Enter the RA
and Dec of your target in
the ÔFieldÕ box, enter the appropriate Equinox.
á
Set the Position
Angle (degrees) to 90 -
set the magnitude limits.
á
Click on Actions
- click on Read GSC
á
Use the left
mouse button to drag the green CCD mosaic so that you place one of the
available guiding CCDÕs on top of a suitable guide star. The R.A. and Dec of the main CCD array
centre will be shown in the Detector box.
(To
obtain data on any star on the chart, place the cursor onto the star and press
the spacebar. The RA, Dec and
Magnitude of the star will be highlighted in the ÔObjects FoundÕ window).
á
On the Autoguider
computer click on the Guide CCDs-WFI bar at the bottom of the screen or on the WFI Guider Selector icon.
á
Click on the
guiding CCD which you placed on top of the guide star in GSFind.
á
Click on the MaxIm
DL – Autoguider Image
bar to return to the autoguider screen.
á
Using the E.T.S.
Motion then Track menu, slew the telescope to the R.A. and Dec
position as shown in the Guide Star Finder ÔDetectorÕ box.
á
Click on Open
Shutter in Cicada (this allows
light to fall onto the Autoguiding CCD without actually taking an exposure).
á
In Maxim DL,
click on the Guide Tab. Set the Exposure
time to 7 seconds,
enter the declination of the guide star (degrees only) into the Declination
Box
á
Click on the Expose button
- click on Start
á
Click the mouse
cursor on the star as displayed in the Autoguider Image window.
WFI
á
Click on the Track button in the Guide Tab
á
Click on Start
- Autoguiding will
commence.
á Close the shutter in Cicada before taking data.
á
In Cicada click
on Open Shutter to pass an
image to the autoguider.
á
Set Exposure
Type to □ Object,
□ Save and Display, set the exposure time.
á
Set the Overscan
Columns to 0 (This
Overscan option is for the 2.3m DBS Astromed controllers. The WFI SDSU controllers have overscan
and bias regions preset).
á
Click on Close
Shutter
** Note:
There is a Shutter Priming Delay **
á
The shutter takes a few seconds to prime, so wait 8-10
seconds after closing the shutter (e.g. after acquiring the guide star) before
exposing.
á
When the shutter is primed and the star
image in the autoguider fades – click on Expose
á
When the image
has finished reading out, click on Open Shutter to re-acquire the guide star.
Check images as they readout in Ximtool
– if the star images are ÒstreakyÓ then you may have to re-boot Cicada to
solve the problem. If the
autoguider image fails to reappear during the exposure – ABORT the
exposure, reset the Run No. and repeat the Data Taking procedure.
(If you have trouble re-booting Cicada, kill all Cicada processes
that are running by typing at the username@monsoon% prompt É ps
–ef | grep cicada
and then the kill %1 or
kill (process
number) -
Remove the ÔcoreÕ file with rm core).

XIMTOOL N
When Ximtool reads out É North is to the right, East is
down.
E
WFI
á Type: deallocate mt1 <Enter>
á
Slew the
telescope to the current Sidereal Time and a Declination of about –31 00
00
á
When the
telescope is in position, at the E.T.S. click on Motion
– click on Halt
AUTOGUIDER
IN THE DOME/TELESCOPE
á
Turn on the dome white
lights.
Observer comment and fault files exist for logging alterations to the instrument and for user comments. WFI users are required to record their experiences in this file for the benefit of others. Access to the file, either to read it or add your comment is available through the MSSSO home page at http://msowww.anu.edu.au/observing/
AUTOGUIDER
SETTINGS (via the GUIDE TAB – ÔSettingsÕ button)
X Axis Cal Time = 1 Backlash = 0
Y Axis Cal Time = 3 Backlash = 0
WFI
Manual Calibration X Speed = 10 Y Speed = 4.83
Check boxesÉ Enable X Axis Enable Y Axis
Aggressiveness = 4
Control Via = Camera Relay
Display Scaling Focal Length = 2000 Display Mode = Pixels Check Delta
Exposure Settings Binning = 1 X Offset = 0 Y Offset = 0 Width = 320
Height = 240
More Autoguider
settingsÉ Guiding Minimum move X = 0.10 Minimum move Y = 0.10
Maximum move X = 3.00
Maximum move Y = 3.00
Delay after correction
(sec) = 0.50
Move Commands Maximum Slew X = 5.00 Maximum Slew Y = 10.00
Delay after commands =
0.10
INITIALISING
CCD HARDWARE
If half of the main CCD
array in Ximtool fails to display, enter the dome and check that the correct
LEDÕs at the rear of the Master and Slave processors are illuminated. The pattern of LEDÕs should be É (G = Green X = Out O = Orange)
Slave Master
O
G
G
X
G
G
G
G
G
G
G
G
X
G
G
O
The Master and Slave
processors are mounted one each side of the WFI dewar. The LEDÕs, which are covered with black
electrical tape, are located on the northern (Control Room) side of the
controllers. Possibly the ÔSlaveÕ
(1) processor will have shut down.
To correct the situation in Cicada under the [Actions] menu, click on Initialise
CCD Hardware
SHUTTER
PRIMING
At large easterly
hour-angles (>-4hrs) the camera shutter can fail to reset. You can recognise this by the star
trails on an image (trails on CCDs 1-4 trail up, on CCDs they 5-8 trail
down). In addition there will be a
ramp in the signal across the CCDs.
Another clue to a failed shutter can be gained by listening to the
shutter operation. There should be
a loud CL-LICK on open and close with the control room door open. A failed shutter sounds different, a
woosy, fainter click due to the fact that only one of the two shutter blades is
operating.
To fix a shutter that fails to reset, you will need to prime
the shutter manually.
Priming the shutter
(Refer to the 40 Inch Telescope Observers Technical Guide
– Exposure Controller – Shutter Prime Procedure).
á
Transfer telescope control to the dome.
á
Move the telescope to ~ -45deg dec, HA=2.5 hrs East.
á
Raise the floor to the level of step 5 on the concrete
stairs
á
On the Wide Field Imager - Exposure Controller,
locate panel E2448 (i.e. the large electronics rack mounted above WFI on the
eastern side of the 40-inch optical tube).
á
Switch the top toggle switch, the Prime Switch,
from Nor (Normal) to the Prime position.
á
Turn to face WFI, at the left of the filter box
you will see a small rectangular inspection port labeled ÔPriming Rod Access
CoverÕ. Remove this cover.
á
To the top/right of the filter box you will see
a knurled knob labeled ÔPriming RodÕ.
Unscrew the rod and insert it into the small round hole at the bottom
right of the Priming Rod Access plate.
á
Gently push in the rod to manually prime the
shutter.
á
When the shutter is primed, withdraw the rod,
replace the cover and rod in its holder.
á
Reset the Prime Switch to Nor (Normal).
You can check that the shutter is primed by
switching the ÔDisplayÕ toggle switch to ÔOnÕ – the ÔReadyÕ LED should
glow – once checked, switch the ÔDisplayÓ toggle switch to ÔOffÕ
To test that the shutter is working:
á
Toggle the ÔModeÕ switch to ÔManualÕ
á
Toggle the ÔDisplayÕ switch to ÔOnÕ
á
Toggle the ÔShutterÕ switch ÔOpenÕ AND then to ÔClosedÕ to
test the shutter.
á
(Don't forget to wait 8 seconds for the shutter to reset
after closing)
á
Return all four
switches UP before resuming observing.
N.B. For normal operation, all four toggle
switches on panel E2448 are in the ÔUPÕ position – Prime=Nor, Mode=Auto, Shutter=Close,
Display=Off
GENERAL
At the
Equator: 4 sec of R.A. = 1Õ arcmin
in Dec or 1 min of R.A. = 15Õ arcmin in Dec
To offset
for non-equatorial declinations = (Declination (in degrees) X Cos) X Offset in Arcmin
WFI chips:
Eight 2k x 2k (2048x4096) CCDs in a 2x4
mosaic
Each sees 13Õx26Õ (arcmin) making a 52Õx52Õ
field
Pixels are 15 micron giving 0.375
arcsec/pixel (on the 40-inch @ f/8)
Gaps between the CCDs are ~20-30 arcseconds
Gains are ~1.5 – 2.1 electrons/ADU
RONs are ~3.5 – 5.5 electrons (To obtain Gain & RON follow the
procedure in the Ò40 Inch Telescope ObserverÕs Technical GuideÓ)
The SDSU controllers read out the entire
array in 55 seconds
The full array image is 140Mb in size -
one CCD is 17.5Mb
The /data/monsoon1 disk on Monsoon holds almost 100Gb – about 4~8 days
data
A DDS3 125m tape holds fractionally over
18Gb compressed (i.e. use allocate
mt1)
Writing to tape takes ~11.5 min/Gb (~3.5
hours per full tape) – donÕt get caught on you last night – SAVE
EARLY!
The plate scale of the WFI is 0.375 arcsec
per pixel, 25 arcsec per mm or 2.5 pixels per sec of arc e.g. if a star image is 5 pixels across
then the seeing is 2 arcsec. The
field of view covers 52 arcminutes with the diagonal being 1.2 degrees.
The values displayed at the lower right
corner of the CICADA_ximtool window show:- The x-axis pixel position/count, the y-axis pixel
position/count and the image/pixel
intensity count. A
general guide to the quality of the
seeing can be made by zooming in on a star and counting the starÕs pixel
width. (5 pixels = 2 arcsecond
seeing). OR Plate Scale X Star FWHM (0.375 X ?)
Pixel size = 10 microns
CCD Array = 320 X 240 pixels
Plate Scale = 0.25 arcsec/pixel or 4 pixels/sec of arc
Field of View = 80 arcsec X 60 arcsec
Pixel size = 15 microns
CCD Array = 8192 X 8192 (Each CCD = 2048 X 4096 plus 20~30 arcsec gaps)
Plate Scale = 0.375 arcsec/pixel or 25 arcsec/mm or 2.5 pixels/sec of arc
Field of View = 52 X 52 arcmin (Diagonal = 1.2 degrees)
DEWAR
VACUUM
The dewar
vacuum rises about 1E-07 per day.
The vacuum is probably Ok up to about 1E-04. LN2 hold time is about 12 hours
CICADA, GIT & Ximtool
General Strategies to deal with CICADA.
Cicada has a tendency to hang, usually during readout, and
sometimes for no obvious reason (even if you are doing none of the bad things
listed below).
The only remedy is to restart it. To do this:
á
In the Observers Window menu select actions - quit-observing-window
á
In the CICADA menu select Start-Observing - Exit_CICADA
From the command line enter:
á
ps -ef | grep cicada
Look for cicada processes. Ignore *_svc.
Kill others with É
Kill –9 (process number) or É
á
Cicada_cleanup -p 0
á
Cicada_cleanup -p 1
á
Cicada_cleanup -p 2
á
Cicada_cleanup -p 3
á
Restart using "cicada &Ó
In general restarting solves Cicada problems, but it's a
pain soÉ
Follow these next practices for relatively trouble-free
operation:-
á
Pause - DON'T use 'pauseÕ. This crashes Cicada so badly you
could need to logout to restart!
á
NumLock - Keep NumLock untoggled. It interferes with ximtool
(and GIT?) operation. In particular, it prevents you using right-mouse
scrolling to change brightness/contrast, and inactivates the ximtool menus.
á
Initialise Hardware - advice is to use this as a last
resort.
Avoid:
á
Using ABORT.
You'll often then need to exit cicada and restart.
á
Doing too much (processor or memory intensive stuff) on
monsoon during readout.
In particular avoid:
á
Beginning or ending a tar command during readout (during
exposing is OK)
á
Doing much in Netscape
á
Copying images (and other operations) in iraf
á
Leaving the little "graphics keystrokes" window
open during readout
á
Changing regions (in the regions window) during readout.
During readout you CAN do these:
á
Change filter, change exposures, change duration
GIT (Graphic & Imaging Tool)
GIT (presently) has a problem with displaying mosaiced
images. It displays them OK on
readout but it cannot set sensible scaling parameters to allow you to calculate
star profiles, get single line plots, etc. The Ximtool window just goes all black or white and displays
a single value.
To overcome these limitations do the following:
á
Untoggle NumLock
á
Select Options -
Start-private-ximtool
á
Select Options – Preferences
In ÒPreferencesÓ É
á
Select category
"General"
á
Untoggle "Mosaic MEF files in Cicada modeÓ
á
Enter the "MEF extension list", i.e. a list of
which CCDs you want available for loading into the private ximtool.
Note that if you are only using CCD numbers 1,2,7,8 (say)
their MEF image extension numbers will be 1,2,3,4. i.e. CCD number and
extension number are different things.
á
Set "Maximum Cicada images" if necessary.
á
The list of extensions you have requested will appear in the
Òimage list" window. Click
one and it loads into the private ximtool.
Now you can get star profiles, cross sections, stats,
etc. But when the next image is read out it will automatically be'loaded into
the private ximtool.
But avoid leaving the little "graphics
keystrokes" window open during readout or GIT will crash out.
![]()
XIMTOOL N
When Ximtool reads out É North is to the right, East is
down.
1 pixel = 0.375arcsec E
To move a star N in
Dec offset on ETS is
NEGATIVE
To Ò Ò S in Dec Ò
Ò
POSITIVE
To Ò Ò W in RA Ò Ò POSITIVE
To Ò Ò E in
RA
Ò
Ò
NEGATIVE
Screen
Preferences for GIT and Ximtool
Select the monitor for these in Options –
Preferences – General.
Common practice puts GIT and Ximtool on screen 1.
Other
Binning
If you change the binning, it may not change back again when
you reset it to 1 x 1, even though the regions window shows it as 1 x 1.
WFI
Watch the file sizes: should be 271440 blocks (full 8xCCDs?)
Bleeding
If readout shows bleeding on brighter stars on some CCDs and
bias changes, restart cicada.
Wrong Frame Size
If there is an error message about ximtool having the wrong
frame size, check that there is not an old cicada ximtool process running:
ps -ef | grep cicada will show it.
Many Red Error Windows
If several red error windows appear from cicada, and
restarting is impossible or doesn't work, try logging out (and back in). That can fix it. If the error messages involve SDSU,
then they probably need a hardware initialisation (from the action menu).
If such error messages are reported when cicada is
restarted, you may need two hwinits.
Do the hwinit immediately after restarting cicada.
COMPUTERS
Mosaic
(in rack) = runs controllers - Monsoon
(on desk) = runs Cicada - Murky (in office) is a bit slow
Data is stored on monsoon in /data/monsoon1/cicada/username
SDSU Controller ÔMasterÕ controls CCDÕs 1,2,3,4 the ÔSlaveÕ controller runs CCDÕs
5,6,7,8
To transfer telescope control to the Dome console:
á
At the ETS PC select Configuration - configure. Toggle Dome Control to ÒmanualÓ.
á
Enter the dome.
á
At the dome console switch the rotary Guide – Slew/Set
knob to ÒSlew/SetÓ
á
Move telescope, dome and floor as necessary
á
Press the large red STOP button if filling dewar or stopping
observing.
To transfer telescope control to the Control Room:
á
Press the "emergency-stop reset" button
(illuminated beside the STOP button)
á
Ensure floor is fully down
á
At the dome console switch the rotary Guide – Slew/Set
knob to ÒAutoÓ
á
á
Go to the control room
á
On the ETS PC select Configuration - configure.
á
Toggle Dome Control to ÒautomaticÓ.
Telescope Tracking
Telescope RA tracking sometimes drops out. Just redo the tracking command in the
ETS computer to restart.
SDSU Cooling
The SDSU controllers are cooled by the NESLAB cooler
(downstairs in corridor through library on the left). If the controllers get hot, this probably means that
the NESLAB has failed. Check the
temperature of the controller heat sinks occasionally - touch them.
IRAF
WFI produces mosaiced images. In iraf you need to specify the extension number in the
filename, as below. If you havenÕt
used IRAF before, refer to
http://msowww.anu.edu.au/computing/lun/lun35/lun35.html
Set the terminal type as xgterm (at the prompt after mkiraf)
and edit login.cl (or loginuser.cl) to set set stdimage = imt4096
set imtype
= ÒfitsÓ
Then:
ximtool & :brings up an ximtool to display images in.
cl
:starts iraf
And:
display filename[#CCD] 1
imcopy filename[#CCD][xmin:xmax,ymin:ymax) outfile.fits
imexamine filename[#CCDI[xmin:xmax,ymin:ymaxl
image operations: v does plot from two cursor input
positions
r gives radial profile of star, fwhm etc.
a gives mag and fwhm etc. for star. no plot
Even for a single CCD readout or a small region readout, you
need the extension filename[1]
Use the mscred package in iraf to reduce mosaiced images.
ds9, saotng are nice alternative display tools. ds9 can
handle mosaics.
TROUBLESHOOTING
If things go wrong - here are some possible causes. Solutions are hopefully in these notes
or somewhere in the sources listed below.
First, follow the suggestions above to keep Cicada as stable as
possible.
SOURCES OF INFORMATION
v
The '40-Inch Telescope (WFI) DIY Manual' (this document)
v
"40 Inch Telescope Observer's Technical GuideÓ
v
http:/lwww.mso.anu.edu.au/observing/40in/DirectIm/40in/
v
http://msowww.anu.edu.aulcomputing/cicada/cicada/
v
The WFI web pages
v
Other hardcopy manuals in the Control Room
v
http://msowww.anu.edu.aulobserving/tc_.manual/
Troubleshooting in the Control Room
My object doesn't appear on Ximtool after readout?
Cloud
Dome is blocking view
Dome shutter blocking view
Camera shutter failure
I see streaks on my image?
Camera shutter failure
Cicada failure – re-boot Cicada
Initialise hardware (a last resort)
I can't analyse images (star profiles, stats, etc) in
Ximtool.
The image just goes to a single value!
Start a private ximtool in GIT.
Cicada has hung?
Exit Cicada, cleanup, restart Cicada.
Stop processor/memory intensive processes
Avoid using abort
Don't use pause
Lots of Red Error Windows, Can't Stop Them!
Kill Cicada from the command line, cleanup and restart
I can't point the telescope more accurately than 20arcsec?
The telescope only points to within about 20aresec -But, use
buttons on round focus/guide/set paddle. Use set or guide speeds. With practice
this becomes fairly accurate (to within a few pixels).
Ximtool menus are inactive
Untoggle NumLock
I can't change brightness/contrast using left mouse
scrolling?
Untoggle NumLock
New images keep loading into my private Ximtool?
Turn off automatic loading of images by toggling the
"cicada mode" button under the GIT Options – Preferences -
General
menu. (This is supposed to work, but sometimes doesn't)
My star images are elliptical
Focus the telescope
Focussing is difficult
Try focusing from same direction to take up backlash
There is bleeding on stars & bias has changed
Restart cicada.
I can't see my guide star (while setting up guiding)
Open the shutter!
I can't find the "Guide CCDs - WFI" chooser
window
Scroll down in the m40slave window
Autoguiding doesn't seem to be working accurately
Set the declination correctly on the "guide" tab
in the Maxim CCD window
Troubleshooting in the Dome
The telescope wonÕt obey the ÒAutoÓ setting of the
dome Auto – Slew/Set
rotary switch
The dome floor is not in the fully lowered position.
The dome doesn't follow telescope?
Check that the dome position encoder (high on the western
wall) is flashing its green LEDs at you. If not - restart ETS (select file -
shutdown, then startup at the prompt).
Put the dome in the ÒautoÓ mode on ETS (select configuration
- configure)
Telescope pointing is bad
Calibrate the pointing (refer to page 4)
Reset the time (Phone 1-1194)
Telescope doesn't move
Ensure floor is fully down
Fault in ETS
(Call technical support)
Telescope moves unexpectedly-by itself
Press large red STOP button on dome console, or on
electronics rack in control room
Shutters don't
open or close
Press the Ôshutter stopÕ button and try again (Listen for
the relays clicking downstairs)
If upper shutter doesn't close it needs to be driven fully
open (onto the external springs) first.
LN2 leaking from filling tube?
Tighten the various nuts on the tube (call technical
support)
LN2 leaking from camera dewar at high ZD angles?
Remove lens-cap before filling camera dewar
Do dome-flats before filling camera dewar
LN2 small tank pressure low?
Small tank needs filling from large tank (call technical
support)
Pressurize small tank with air line by connecting air line
to top of LN2 tank.
SDSU controllers feel hot
Water cooler has failed. Check it and restart it, if necessary.
Dewar vacuum rising or focal plane temperature not
holding
Pump dewar (call technical support)
Ver. 02/11/05