2·3m Telescope Control for DBS Observers


Contents

Dome lights
Telescope pointing calibration
Rotator selection and control
Using the Nasmyth platform control handset
The telescope CONFIGURE command
The telescope SWITCH command

This page contains descriptions of some telescope controls that are of particular interest to users of the DBS. It is not an exhaustive user manual for the telescope as a whole. Further details of the telescope control system may be found in the MSSSO Telescopes Command Reference Manual, a hardcopy version of which are stored in the console room.


Dome lights

Lights within the building are controlled by two lighting switches at the top centre of the console. To switch off all lights (except the flat-field illumination and arcs!) set to FLUOROS OFF and INCAND OFF. There is also a dial to control brightness of the flat-field illumination lamps. Always restore the lights when you have finished observing and leave the building.


Telescope pointing calibration

Because this telescope is an Alt-Azimuth telescope, good tracking performance depends upon good pointing calibration, and this pointing calibration is more involved than for an Equatorial telescope. In particular, pointing calibration must be carried out with the star centred in the current aperture (i.e., you must know which point on the screen corresponds to the current aperture, and the star must be placed there.) Telescope apertures represent positions in the focal plane, and do not relate to real instrument apertures unless you define them to do so. Only one such relationship has been pre-defined; aperture 1 on the Double Beam Spectrograph [DBS]. On other instruments you must define or calibrate an aperture in order to successfully calibrate the telescope pointing. Any errors in calibration will lead to the object drifting (in an unpredictable direction) throughout observations. For users of the DBS, good calibration can be achieved by following these simple steps:
  • Select a suitable bright star (mag 6 - 8) to use for pointing calibration. It should be at zenith distance greater than 20° to minimise field rotation during the calibration.
  • Send the telescope to the star's position. e.g.,
    TEL$ TRACK/COORD 01 02 03.45 -50 40 32.1 J2000
    The star should then be visible in a short exposure using the Apogee autoguider camera. If the star is not visible in the guider, check the suggestions on the trouble shooting page.
  • Select aperture 1. This aperture has been pre-defined to be the intersection of the slit and the tip of the decker.
    TEL$ APERTURE 1
  • Move the telescope to place the star at this position. This is easiest achieved using the autoguider "move-to" command.
  • Calibrate the pointing here.
    TEL$ CALIBRATE POINTING

    If you wish to perform observations with the target away from this point, then proceed to follow these steps:

  • Move the star to the desired place.
  • Define this place as an aperture to the telescope system , giving it a 1 or 2 character name (in this example-"xy").
    TEL$ APERTURE/HERE xy
    Aperture xy is now defined and automatically selected as the current aperture. Future CALIBRATE POINTING commands may now be executed at aperture xy or any other properly defined aperture, so long as that aperture is selected as the current aperture first.

    If you have any problems with the above procedure, you might try defining your own new aperture from scratch. This is a slightly more involved procedure requiring you to re-acquire the same object at either end of the slit and with the rotator at different orientations. For further details see CALIBRATE APERTURE in the MSSSO Telescopes Command Reference Manual.


    Rotator selection and control

    Rotator selection

    Moving the rotator

    Modes of operation

    Selecting rotator orientation


    Control of telescope from the rotator handset

    Note: If you need to drive the rotator a long way with the handset, you can ensure that you do not go beyond its safe limits by first reseting the rotator angle to zero from the TCS console. Either use the rotator control buttons on the console or type
    TEL$ CONFIGURE ROTATOR_REFERENCE STATIONARY
    TEL$ ROTATOR 0

    In addition to control of the telescope from the console, control is also available at the telescope focal station by means of the rotator handset. There are three handsets, one each for the rotators at the three telescope foci. Only the handset at the currently selected rotator may be used to control the telescope motion and the focusser; the other two provide only manual control of their respective rotators.

    Image of rotator handset.
    The pushbutton/annunciator POWER (lamptest) at the top of the handset is illuminated whenever the rotator is ON in manual or automatic modes. It can also be pressed at any time to test all of the annunciator lamps or to see whether TELESCOPE SYSTEMS have been switched on. A brightness control knob at the bottom is used to set the correct level of illumination for all the annunciators on the handset.

    At the top of the handset two pushbuttons give control of the mode of operation of the rotator:

    Pressing the blue MANUAL pushbutton brings the rotator on in manual mode (provided TELESCOPE SYSTEMS have been switched on) and after a short delay ROTATOR READY is signalled by the green annunciator at the top right. The rotator may now be controlled manually as described below.

    Pressing the green AUTO/OFF pushbutton brings the rotator on in automatic mode (provided TELESCOPE SYSTEMS have been switched on and the rotator in question has been selected). After a short delay ROTATOR READY is signalled by the green annunciator at the top right and the rotator is now controllable from the computer. If that rotator was not the one selected, it simply switches off.

    A group of annunciators down the left-hand side of the handset indicate the state of the rotator and telescope system hardware.

    To actually move the rotator, make sure the handset is in MANUAL mode, hold down the CONTROL button and press ROTATOR CW or ROTATOR CCW for clockwise and counter-clockwise rotation respectively. When finished return the rotator to automated control by pressing AUTO/OFF.

    Observers should avoid repeated switching between automatic and manual modes whilst the telescope is tracking, because it will drop out of the tracking state and it is possible for large impulses to be imparted to the rotator drive gears. If you switch to manual at the handset whilst the telescope is tracking, allow about 10 seconds before switching back to automatic so that the control program has time to halt the rotator smoothly.


    The telescope CONFIGURE command

    See also the Command Reference Manual.

    The majority of telescope system logical configuration changes are made via the CONFIGURE command. Physical movement of actual telescope hardware is mostly controlled via the SWITCH command. The syntax for CONFIGURE is simply
    TEL$ CONFIGURE variable-name value

    The configuration variables and their possible values are listed below:
    Configuration variablePossible values
    Observer_namestring variable set at login time only
    CFflefile specification for telescope control file
    Instrument_identstring
    SecondaryNasmyth, IR_Cassegrain
    Focal_StationNasmyth_A, Nasmyth_B, Cassegrain, Stowed
    Focus_ControlManual, Automatic, Compensated
    Windscreen ControlOpen, Closed, Tracking, Vertical_Only Tracking, Off
    Windscreen_ApertureNormal, Wide_for_Finder
    Windscreen_Normal-Clearance(numeric value in metres)
    Default_File_Equinoxstring must be valid system equinox specification
    Tracking_Equinoxstring must be valid system equinox specification or the keyword FILE
    Default_Temperaturetemperature in degrees Celsius
    Default_Pressureatmospheric pressure in mBar
    Default_Humidityrelative humidity in %
    Effective_Wavelengtheffective wavelength for detector and source in nm
    Rotator_ReferenceNot_In_Use, Position_Angle, Vertical_Angle, Stationary
    Rotator_OrientationSlit, Radial, Apertures, Angle
    Rotator_Ap1aperture ident for first aperture
    Rotator_Ap2aperture ident for second aperture
    Rotator_Orientation_Angleangle in degrees
    Rotator_CCW_Limitangle in degrees (software limit to rotator motion)
    Rotator_CW_Limitangle in degrees (software limit to rotator motion)


    The telescope SWITCH command

    See also the Command Reference Manual.

    Physical movement of actual telescope hardware is mostly controlled via the SWITCH command. Changes to system configuration choices are made via the CONFIGURE command, though the division of duties between SWITCH and CONFIGURE is not always obvious. The syntax for SWITCH is simply
    TEL$ SWITCH variable-name value

    The switch variables and their possible values are listed below:
    Switch variablePossible values
    Telescope_SystemsOn, Off, Reset
    Ventilation_FansOn, Off
    Shutter* Closed
    Mirror_CoverOpen, Closed
    Flatfield_IlluminationOn, Off

    * The value `Open' also exists, but is available to the TCS only. It cannot be set from the user terminal for safety reasons.

    Note: Flatfield_Illumination is the only system for which there is no physical control button on the console, and thus the only one you are likely to use.


    Last modified: August 10, 2005