Observers Guide


Details of the telescope control system may be found in the MSSSO Telescopes Command Reference Manual, a hardcopy version of which is stored in the console room.

Getting going quickly

A brief crib sheet covering most of what you have to do every night. It is principally intended for those who have used the telescope before. Detailed descriptions of how to run the instrument are given below.


Instrument changes

Observatory Staff attach the correct instrument and computers to drive that instrument. Talk to them about which computers drive your instrument, and discuss any recent changes. Grating and dichroic changes, focus etc are the observers' responsibility. New and unexperienced observers must make arrangements to have someone familiarise them with the instrument well in advance of their scheduled run. Read the "Beginning Your Run" section of the Getting going quickly manual.


Starting up the telescope

To ready the telescope for observations the equipment comprising TELESCOPE SYSTEMS must be switched on and the group of computer processes which comprise the TELESCOPE CONTROL PROGRAM must be installed and started.
  • Press CONSOLE ON on the PLANT CONTROL PANEL (upper left of console). Various indicators will now light on the PLANT CONTROL PANEL and the VDUs will come on. When the console is off, TELESCOPE SYSTEMS cannot be switched on and nor will the TELESCOPE CONTROL PROGRAM function correctly.
  • Press TELESCOPE SYSTEMS ON on the PLANT CONTROL PANEL. Three green indicators BUILDING DRIVE READY, AUXILIARY SYSTEMS READY and SERVOS READY should appear. A red FAULT indicator or less than three green READY lamps indicates a malfunction which will prevent you observing. See the trouble shooting guide. At this stage one can, if desired, open the building shutter, the primary mirror cover and start the ventilation fans. Always ensure that the mirror cover is fully closed before opening or closing the building shutter. This will prevent any detritus from the shutters falling onto the mirror.
  • The telescope control computer is called MAIA.
  • To log on, press BREAK at the telescope control keyboard.
    Username is TELESCOPE.
    For password, see the technical staff.
    Unless you have your own account on MAIA, enter VISITOR when prompted "Please enter Observer's identification".
  • Type STARTUP at the $ prompt and wait until the TEL$ prompt appears.
  • The current setup can be shown by entering
    TEL$ DISPLAY CONFIGURATION
    To return to the observer's display, use the command
    TEL$ DISPLAY
  • If you have your own account on MAIA, load your telescope control file using the command TEL$ CFILE your-filename
    Otherwise, the setup may be changed by entering individual configuration commands, e.g.,
    TEL$ CONFIGURE FOCAL_STATION NASMYTH_A
    TEL$ CONFIGURE FOCUS_CONTROL COMPENSATED
    TEL$ CONFIGURE ROTATOR_REFERENCE POSITION_ANGLE
    TEL$ CONFIGURE ROTATOR_ORIENTATION SLIT
    See MSSSO Telescopes Command Reference Manual, a hardcopy of which is stored in the console room.
  • Open communications with DBS autoguider hardware by typing
    TEL$ ENLIST TXA6
  • Lights within the building are controlled by two lighting switches at the top centre of the console. Set to FLUOROS OFF and INCAND OFF (or INCAND PRESET).
  • Open the building shutter. Press BUILDING SHUTTER OPEN on the console.
  • To start the building ventilation fans, press VENTILATION FANS ON on the console.
  • Open the mirror cover. MIRROR COVER OPEN on the console.
  • Calibrate telescope pointing.
  • Focus the telescope by using the autoguider camera as a TV display while driving the focus controls manually. See the autoguider manual sections 5 & 6 for details.


    Taking exposures / The CCD controllers

    VERY IMPORTANT:
    There have been big changes to the CCD controllers. See this page before reading any further.

    The DBS cameras are currently run by Astromed controllers connected to a pair of workstations running UNIX, located on the left of the Telescope Control System. The second one on the left of the TCS is called "mouldy". From there the Red arm of the DBS, and the Imager, is controlled. The DBS Blue is run from "moist", the third one on the left of the TCS. This is a temporary arrangement and you may need to see the staff to make sure the computer is running the correct instrument, or to learn how to swap between DBS Red and Imager. The DBS Blue data is usually stored on the workstation "mallorn" in the Peripherals room. You can log on it from the workstation "moist", located next to "mouldy". Running "Teldisk" on CICADA at the beginning of your run will automatically select the right output path for your images.

    The Astromed's command structure is rather obscure, so the CCDs are controlled via a collection of macros which have been predefined and are loaded automatically on the CICADA interface. To start, type "cicada &" in any window of the appropriate workstation and select the DBS Blue on "moist" and the DBS Red on "mouldy" (if unsure, check with Staff to learn the names of the CCDs you should select). Set the data frame number using CICADA. Numbers of up to four digits may be used.

    In order that the correct header information be read from telescope control computer, the exposure must be started on the red arm of the spectrograph before starting the blue exposure. IMPORTANT: the DBS-Imager setup PC does not interact with CICADA, therefore whatever information appears there (slit width, filters, lamps used) must be entered in the header MANUALLY using CICADA, or recorded in your personal observing logbook/logfile.

    At the end of the night, CCDs should be left on "CCD temp + flush". Select the appropriate macro from the CICADA pulldown menu.

    The controllers are initialised using the "Initialise CCD Hardware" button on the CICADA menu. This should be done at the beginning of the run and every time the bias level counts are abnormal.


    DBS Hardware control

    Computer control of DBS hardware such as slit width, decker and arc lamps is from the PC labelled "Imager-DBS" at the very left hand end of the main console. Normally the control software will already be running. If the DOS prompt is showing, run dbs from the \dbs directory. i.e.,
    cd \dbs
    dbs
    Then select "control -> status" from pull-down menus.

    The mouse driven menus should be largely self-explanatory, but some details are given below.

    The decker
    To move the decker, click on "Decker" in the window labelled "M2" to open the control dialogue box. It cannot be set to specific positions, but rather is manually driven into (using button labelled ">>") and out from (using "<<") the slit. With the dome lights or one of the arc lamps (the Quartz lamps are too bright) on, the decker is clearly visible in the auto-guider camera. Use this to approximately position the decker or to see when you have driven the decker clear of the slit without having to take an exposure.

    Setting the slit width
    To change slit width, click on "Slit", located above "Decker" in the "M2" window to open the control dialogue box. Either specify the required value (measured in arc-seconds) in the dialogue box and click "Apply" or drive the slit manually using the buttons labelled ">>" and "<<".

    You can check the slit visually on the autoguider camera using scattered light in the dome with the arc mirror out. The orientation is reversed - e.g., North up, East to right. When slit is in position angle 90, the left end of the slit points East.

    Changing filters

    Selecting arc lamps
    NeAr, CuAr, FeAr and CuHe arc lamps and two quartz flat-fields are available. Only one lamp may be selected at a time and it is specified in the "Arcs" dialogue in the "M1" window. A mirror folds the light from the lamps into the spectrograph and is controlled from the "ArcMirror" dialogue, also in the "M1" window. The lamps are switched on automatically when the arc mirror is "IN". If the arc lamp is too bright, neutral density filters are available. See Changing filters above.


    Taking calibration data

    Bias Frames

    The repeatable bias structure is small; typically ±1 in B and ±3 in R. The bias level is of order 1000, so simple consideration of the Poisson statistics shows that you need >>10 bias frames to really measure it properly. Do not mistake random noise which often appears for genuine bias structure. Using too few biases which include random rubbish will make you data worse, not better. The CCD controller sometimes creates considerable noise in the form of coarse horizontal striping, which cannot be corrected by bias frames (it looks a little like a badly tuned TV signal.) This must be cleared by running "Initialise CCD Hardware" from the Cicada pull-down menus, but this also resets the bias levels rendering the old bias frames unreliable.


    An example subsection of an R bias frame.

    Arc lamps for wavelength calibration

    NeAr, CuAr, FeAr and CuHe arc lamps are available. The CuHe lamp actually appears to be CuHeAr. The blue end of the spectrum, is certainly dominated by Argon lines. To switch arcs on, see Selecting arc lamps above.

    When choosing an appropriate exposure time, it is helpful to run `Autoscale and Update Display' (macro 15 on menu tablet) and then make a horizontal cross section plot through the arc image (macro 5) to ensure the stronger lines are not saturated.

    Sample exposures with 2arcsec slit
    300B 600B 1200B(I) 1200B(II) 158R* 316R* 600R 1200R
    NeAr 10 30 1 1
    CuAr 10 30 1 1
    FeAr 10 30 1 1
    CuHe 10 30 1 1
    Quartz 1 3 1
    Quartz 2 3 1

    * Note: When using the low resolution gratings, the arc lamps are likely to be too bright for even a 1 second exposure. You will need to use the arc lamp neutral density filter. Starting from the fully transparent position, (i.e., display reading `Rev Limit'), drive the filter in using the ">" button for about 4 - 6 seconds. This should be sufficient to allow a 1 second exposure without saturation. You will also have to increase the blue arm exposure to compensate. The QI2 lamp is slightly fainter than QI1, so may be better if you find your flats saturating at 1 second.

    You may also find the DBS arc atlas useful.

    Taking flat fields (Quartz lamps, Dome flats, Twilights)

    Several types of flat field image are possible. Two internal quartz lamps are available in place of the normal comparison arcs. Some observers have reported finding dome flats more effective except in the UV. The new red detector has severe fringing that can be removed with a flat field frame taken at the same spectrograph orientation as for the star.

    Quartz lamps: Two separate, but similar lamps are available in the arc lamp assembly. They are selected through the DBS control computer. Without using filters they are rather bright and can saturate the red CCD in less than one second at low dispersions. The QI2 lamp is slightly fainter than QI1. The blue arm will require longer exposures.

    Dome Flats: A ring of small lamps is installed on the telescope top-end ring and may be used to illuminate the dome ceiling. The lamps are controlled from the telescope console with the commands:
    TEL$ SWITCH FLAT ON
    and
    TEL$ SWITCH FLAT OFF
    The following suggested procedure requires you to first start the telescope control computers and log on. The fluorescent strip lights in the dome must be switched off. The incandescent lights are less important. Move the telescope to zenith park and then slew it down slightly in altitude to prevent any harsh specular reflection off the ceiling. For example:
    TEL$ PARK
    TEL$ SLEW * 75
    Flats are best taken against the wind-screens rather than the actual dome roof, so issue the command:
    TEL$ CONFIGURE WINDSCREEN_CONTROL CLOSED
    Open the mirror covers once the wind-screens have stopped moving. The lamps' brightness is controlled by a black round dial marked "Flat field illumination control" on the telescope console. It is located under the other light controls near the top centre of the console. With the lamp brightness set at around three-quarters, expect exposures of between 10 and 100 seconds, depending on dispersion. The blue arm typically requires about five times the exposure of the red.

    Twilight flats: The illumination pattern across the slit is normally different for flat fields than data frames taken against a dark night sky. Twilight fields can be very effective at correcting the flats before they applied to data frames. See, for example, "A User's Guide to Reducing Slit Spectra with IRAF" by Massey, Valdes & Barnes.


    End of night procedures

    Shutting down the telescope at the end of the night is largely automated.
  • At the telescope control terminal, enter
    TEL$ PARK
  • When the telescope stops moving, enter
    TEL$ SHUTDOWN
    and wait until the system advises that it is safe to leave. The mirror cover and building shutter will be closed automatically, but should be checked by the observer.
  • Finally you may optionally enter
    TEL$ CONSOLE OFF
  • Restore building lights to normal.
  • Remember to fill dewars.


    This page is http://msowww.anu.edu.au/observing/2.3m/DBS/dbs_observing.html

    Last modified: Tuesday January 03, 2005