AUSTRALIAN
NATIONAL UNIVERSITY System Design Note 4.09 Created: 24 April 2002 Last modified: 4 May 2002 |
ACCEPTANCE TEST PLAN
Jan van Harmelen
Research School of Astronomy
and Astrophysics
Institute of Advanced
Studies
Australian National
University
Revision History
|
Revision No. |
Author & Date |
Approval & Date |
Description |
|
Revision 1 |
Jan van Harmelen 22 January 2001 |
Peter J. McGregor 01 March 2001 |
CDR version. |
|
Revision 2 |
Peter J. McGregor 11 September 2001 |
Jan van Harmelen 03 May 2002 |
Revised after CDR. |
Contents
10 Electrical and
Electronic Requirements
16 Summary of Test
Equipment and Test Software
The Acceptance Test Plan describes the inspections, demonstrations, and tests leading to acceptance of the Near-infrared Integral Field Spectrograph, consistent with the requirements of the instrument.
The required contents of the Acceptance Test Plan are described in clause 4.3(b)(10) of Exhibit A, “Statement of Work”, of the NIFS Design and Fabrication Contract.
|
Document
ID |
Source |
Title |
|
AURA |
NIFS Design and Fabrication Contract |
|
|
AURA |
NIFS Design and Fabrication Contract, Statement of Work |
|
|
RSAA |
NIFS Operational Concept Definition Document |
|
|
RSAA |
NIFS Functional and Performance Requirements Document |
|
|
RSAA |
NIFS Flexure Test Plan |
|
|
RSAA |
Structural System Modifications |
This document describes the acceptance test plan for the Gemini Near-infrared Integral Field Spectrograph (NIFS). It contains the following components:
· A summary matrix listing all of the requirements with an indication of how each requirement will be verified.
· Descriptions of the procedures for each inspection, demonstration, and test to be performed in sufficient detail to allow AURA to evaluate the appropriateness of each procedure.
· Descriptions of the inspections, demonstrations, and tests to be performed after arrival at the delivery location to confirm that NIFS continues to meet the specifications after transportation.
· Descriptions of the facilities and resources that will be needed at the delivery location to support assembly and testing of the instrument after it has been delivered.
The matrix below relates each of the requirements listed in the NIFS Functional and Performance Requirements Document (FPRD; SDN0003.02) to a method of verifying compliance described elsewhere in this document. The NIFS FPRD includes all requirements listed in the NIFS Operational Concept Definition Document (OCDD; SDN0003.01). This in turn includes the requirements stated in the NIFS Design and Fabrication Contract.
Notes:
1. Parts of NIFS are built to designs by the University of Hawaii (UH) and parts of the OIWFS are manufactured at UH. These designs have been accepted without analysis by the NIFS developers.
|
Requirement |
Description |
Acceptance Test |
||
|
Section |
Pre-Delivery |
Post-Delivery |
||
|
REQ-OCD-0001 |
Operation with ALTAIR |
5.1 |
… |
Telescope test |
|
REQ-OCD-0002 |
Wavelength Coverage |
5.2 |
Laboratory test |
Telescope test |
|
REQ-OCD-0003 |
Spatial Resolution |
5.3 |
Laboratory test |
Telescope Test |
|
REQ-OCD-0004 |
Field of View, Distortion |
5.4 |
Laboratory test |
… |
|
REQ-OCD-0005 |
Strehl Ratio |
5.5 |
Laboratory test |
… |
|
REQ-OCD-0006 |
Spectral Resolution |
5.6 |
Laboratory test |
… |
|
REQ-OCD-0007 |
System Efficiency |
5.7 |
… |
Telescope test |
|
REQ-OCD-0008 |
System Emissivity |
5.8 |
Laboratory test |
Telescope test |
|
REQ-OCD-0009 |
Scattered Light |
5.9 |
Laboratory test |
… |
|
REQ-OCD-0010 |
Ghost Images |
5.10 |
Laboratory test |
… |
|
REQ-OCD-0011 |
Contrast Ratio |
5.11 |
Laboratory test |
Telescope test |
|
REQ-OCD-0012 |
Detector Read Noise |
5.12 |
Laboratory test |
Telescope test |
|
REQ-OCD-0013 |
Detector Dark Current |
5.13 |
Laboratory test |
Telescope test |
|
REQ-OCD-0014a |
OIWFS for Tip-tilt and Focus |
5.14.1 |
Laboratory test |
Base Facility test |
|
REQ-OCD-0014b |
OIWFS Field of View |
5.14.2 |
Laboratory test |
… |
|
REQ-OCD-0014c |
OIWFS Offsetting Accuracy |
5.14.3 |
Laboratory test |
… |
|
REQ-OCD-0014d |
OIWFS Position Repeatability |
5.14.4 |
Laboratory test |
… |
|
REQ-OCD-0014e |
OIWFS Sensitivity |
5.14.5 |
… |
Telescope test |
|
REQ-OCD-0015 |
Mechanisms Time to Function |
5.15 |
Laboratory test |
… |
|
REQ-OCD-0016 |
Not used |
- |
- |
- |
|
REQ-OCD-0017 |
Downtime |
5.16 |
… |
… |
|
REQ-OCD-0018 |
Data Assessment |
5.17 |
Demonstration |
… |
|
REQ-OCD-0018a |
View and Observe Mode QLDs |
5.17.1 |
Demonstration |
… |
|
REQ-OCD-0018b |
View Mode data displayed |
5.17.2 |
Demonstration |
… |
|
REQ-OCD-0018c |
Observe Mode data displayed and archived |
5.17.3 |
Demonstration |
… |
|
REQ-OCD-0018d |
QLDs display raw data, spectrally compressed data and spectrum |
5.17.4 |
Demonstration |
… |
|
REQ-OCD-0018e |
QLD frame subtraction |
5.17.5 |
Demonstration |
… |
|
REQ-OCD-0018f |
Specification of spectral compression |
5.17.6 |
Demonstration |
… |
|
REQ-OCD-0018g |
Continuum subtraction |
5.17.7 |
Demonstration |
… |
|
REQ-OCD-0018h |
Continuum subtraction specification |
5.17.8 |
Demonstration |
… |
|
REQ-OCD-0018i |
Effect of Flip Mirror on QLDs |
5.17.9 |
Demonstration |
… |
|
REQ-OCD-0018j |
Spectra display specification |
5.17.10 |
Demonstration |
… |
|
REQ-OCD-0018k |
View Mode parameters |
5.17.11 |
Demonstration |
… |
|
REQ-OCD-0018l |
Observe Mode parameters |
5.17.12 |
Demonstration |
… |
|
REQ-OCD-0018m |
Display Time |
5.17.13 |
Laboratory test |
… |
|
REQ-OCD-0019a |
Focal Plane Mask Wheel Blank |
5.18.1 |
Laboratory test |
… |
|
REQ-OCD-0019b |
Filter Wheel Blank |
5.18.2 |
Laboratory test |
… |
|
REQ-OCD-0019c |
Focal Plane Wheel Slits |
5.18.3 |
Laboratory test |
… |
|
REQ-OCD-0019d |
Artificial Calibration Sources |
5.18.4 |
… |
Telescope test |
|
REQ-OCD-0020 |
Polarimetry |
5.19 |
Laboratory test |
… |
|
REQ-OCD-0021 |
System Sensitivity |
5.20 |
… |
Telescope test |
|
REQ-OCD-0022 |
Point sources K≥3 |
5.21 |
… |
Telescope test |
|
REQ-FPR-0001 |
Not used |
- |
- |
- |
|
REQ-FPR-0002 |
Cold Stop Alignment |
6.1 |
… |
Telescope test |
|
REQ-FPR-0003 |
Focus |
6.2 |
Laboratory test |
… |
|
REQ-FPR-0004 |
Not Used |
- |
- |
- |
|
REQ-FPR-0005 |
OIWFS Feed |
6.3 |
Inspection |
… |
|
REQ-FPR-0006 |
Not Used |
- |
- |
- |
|
REQ-FPR-0007 |
OIWFS Filters |
6.4 |
Laboratory test |
… |
|
REQ-FPR-0008 |
OIWFS Wavelength Sensitivity |
6.5 |
Reference |
… |
|
REQ-FPR-0009 |
OIWFS Focus |
6.6 |
Laboratory test |
… |
|
REQ-FPR-0010 |
Optical Baffling |
6.7 |
Reference |
… |
|
REQ-FPR-0011 |
Not Used |
- |
- |
- |
|
REQ-FPR-0012 |
Dust Cover |
6.8 |
Demonstration |
… |
|
REQ-FPR-0013 |
Dust Removal |
6.9 |
Demonstration |
… |
|
REQ-FPR-0014 |
Vignetting |
6.10 |
Reference |
… |
|
REQ-FPR-0015 |
Order Sorting Filters |
6.11 |
Reference |
… |
|
REQ-FPR-0016 |
Not Used |
- |
- |
- |
|
REQ-FPR-0017 |
Cold Stop |
6.12 |
Reference |
… |
|
REQ-FPR-0018 |
Coatings Characteristics |
6.13 |
Reference |
… |
|
REQ-FPR-0019 |
Coatings and Thermal Cycling |
6.14 |
Reference |
… |
|
REQ-FPR-0020 |
Not Used |
- |
- |
- |
|
REQ-FPR-0021 |
Optical Components in Vacuum |
6.15 |
Reference |
… |
|
REQ-FPR-0022 |
Optical Comp Temp Cycling |
6.16 |
Reference |
… |
|
REQ-FPR-0100 |
Detector Bias Variations |
7.1 |
Laboratory test |
Telescope test |
|
REQ-FPR-0101 |
Detector Gain Variations |
7.2 |
Laboratory test |
Telescope test |
|
REQ-FPR-0102 |
Detector Format |
7.3 |
Reference |
… |
|
REQ-FPR-0103 |
Not Used |
- |
- |
- |
|
REQ-FPR-0104 |
Detector Characteristics |
7.4 |
Reference |
… |
|
REQ-FPR-0105… REQ-FPR-0114 |
Not used |
- |
- |
- |
|
REQ-FPR-0115 |
Detector Mechanical Interface |
7.5 |
Reference |
… |
|
REQ-FPR-0116 |
Detector Thermal Interface |
7.6 |
Reference |
… |
|
REQ-FPR-0117 |
Detector Optical Interface |
7.7 |
Laboratory test |
… |
|
REQ-FPR-0118 |
Detector Electrical Interface |
7.8 |
Reference |
… |
|
REQ-FPR-0119 |
Detector Controller |
7.9 |
Inspection |
… |
|
REQ-FPR-0120 |
Detector Controller Mech I/F |
7.10 |
Inspection |
… |
|
REQ-FPR-0121 |
Detector Controller Thermal I/F |
7.11 |
Inspection |
… |
|
REQ-FPR-0122 |
OIWFS Detector |
7.12 |
Reference |
… |
|
REQ-FPR-0123 |
OIWFS Detector Controller |
7.13 |
Inspection |
… |
|
REQ-FPR-0124 |
OIWFS Det Contr Mech I/F |
7.14 |
Inspection |
… |
|
REQ-FPR-0125 |
OIWFS Det Contr Thermal I/F |
7.15 |
Inspection |
… |
|
REQ-FPR-0126 |
OIWFS Detector Optical I/F |
7.16 |
Reference |
… |
|
REQ-FPR-0127 |
OIWFS Detector Electrical I/F |
7.17 |
Reference |
… |
|
REQ-FPR-0200 |
Cold Stop Alignment |
8.1 |
Laboratory test |
… |
|
REQ-FPR-0201 |
Movement of Spectra |
8.2 |
Laboratory test |
Telescope test |
|
REQ-FPR-0202 |
Tracking with OIWFS |
8.3 |
Reference |
… |
|
REQ-FPR-0203 |
Not Used |
- |
- |
- |
|
REQ-FPR-0204 |
Mech and Thermal Tolerances |
8.4 |
Reference |
… |
|
REQ-FPR-0205 |
Temperature Gradients |
8.5 |
Reference |
… |
|
REQ-FPR-0206 |
Thermal Transients |
8.6 |
Reference |
… |
|
REQ-FPR-0207 |
Space Requirements |
8.7 |
Inspection |
… |
|
REQ-FPR-0208 |
Electronic Enclosures |
8.8 |
Inspection |
… |
|
REQ-FPR-0209 |
Access to Electr Enclosures |
8.9 |
Inspection |
… |
|
REQ-FPR-0210 |
Access to Vacuum Ports |
8.10 |
Inspection |
… |
|
REQ-FPR-0211 |
Access to Cooling Water Ports |
8.11 |
Inspection |
… |
|
REQ-FPR-0212 |
Access to Dry Air Ports |
8.12 |
Inspection |
… |
|
REQ-FPR-0213 |
Mechanical Connections |
8.13 |
Inspection |
… |
|
REQ-FPR-0214 |
Total Mass |
8.14 |
Laboratory test |
… |
|
REQ-FPR-0215 |
Center of Gravity |
8.15 |
Laboratory test |
… |
|
REQ-FPR-0216 |
Balance Tolerance |
8.16 |
Reference |
… |
|
REQ-FPR-0217 |
Ballast Weight |
8.17 |
Inspection |
… |
|
REQ-FPR-0218 |
Spectrograph Thermal Stability |
8.18 |
Laboratory test |
… |
|
REQ-FPR-0219 |
Cool Down Time |
8.19 |
Laboratory test |
… |
|
REQ-FPR-0220 |
Warm Up Time |
8.20 |
Laboratory test |
… |
|
REQ-FPR-0221 |
Detector Thermal Stability |
8.21 |
Laboratory test |
… |
|
REQ-FPR-0222 |
Vibration |
8.22 |
Reference |
… |
|
REQ-FPR-0223 |
Vacuum System Holding Areas |
8.23 |
Inspection |
… |
|
REQ-FPR-0224 |
Vacuum Pump Capacity |
8.24 |
Inspection |
… |
|
REQ-FPR-0225 |
Vacuum System Operating Proc |
8.25 |
Inspection |
… |
|
REQ-FPR-0226 |
Vacuum System Test Set-up |
8.26 |
Inspection |
… |
|
REQ-FPR-0227 |
Mechanism Safety |
8.27 |
Laboratory test |
… |
|
REQ-FPR-0228 |
Metric Dimensioning |
8.28 |
Inspection |
… |
|
REQ-FPR-0229 |
Metric Dimensions on Drawings |
8.29 |
Inspection |
… |
|
REQ-FPR-0230 |
Metric Fasteners |
8.30 |
Inspection |
… |
|
REQ-FPR-0231 |
Repeatability of Configuration |
8.31 |
Laboratory test |
… |
|
REQ-FPR-0232 |
Instrument Handling |
8.32 |
Inspection |
… |
|
REQ-FPR-0300 |
Control System Operability |
9.1 |
Demonstration |
… |
|
REQ-FPR-0301 |
Configuration Time |
9.2 |
Reference |
… |
|
REQ-FPR-0302 |
Impact on Mechanism Accuracy |
9.3 |
Laboratory test |
… |
|
REQ-FPR-0303 |
Impact on Science Performance |
9.4 |
Laboratory test |
… |
|
REQ-FPR-0304 |
Temperature Control |
9.5 |
Demonstration |
… |
|
REQ-FPR-0305 |
Detector Temperature |
9.6 |
Reference |
… |
|
REQ-FPR-0306 |
Optical Elements Temperature |
9.7 |
Reference |
… |
|
REQ-FPR-0307 |
Rate of Temperature Change |
9.8 |
Laboratory test |
… |
|
REQ-FPR-0308 |
Speeding the Warming up |
9.9 |
Laboratory test |
… |
|
REQ-FPR-0400 |
Not Used |
- |
- |
- |
|
REQ-FPR-0401 |
Grounding and Shielding |
10.1 |
Reference |
… |
|
REQ-FPR-0402 |
Electrostatic Discharge |
10.2 |
Reference |
… |
|
REQ-FPR-0403 |
Not Used |
- |
- |
- |
|
REQ-FPR-0404 |
Not Used |
- |
- |
- |
|
REQ-FPR-0405 |
Surface Temperatures |
10.3 |
Laboratory test |
… |
|
REQ-FPR-0406 |
Not Used |
- |
- |
- |
|
REQ-FPR-0407 |
Temperature Monitoring |
10.4 |
Demonstration |
… |
|
REQ-FPR-0408 |
Temperature Sensor Locations |
10.5 |
Inspection |
… |
|
REQ-FPR-0409 |
Temperature Sensor Interfaces |
10.6 |
Demonstration |
… |
|
REQ-FPR-0500 |
Software Design Requirements |
11.1 |
Laboratory test |
Base Facility test |
|
REQ-FPR-0501 |
Standard Configuration |
11.2 |
Inspection |
… |
|
REQ-FPR-0502 |
EPICS System |
11.3 |
Inspection |
… |
|
REQ-FPR-0503 |
Use of CICS |
11.4 |
Inspection |
… |
|
REQ-FPR-0504 |
Not Used |
- |
- |
- |
|
REQ-FPR-0505 |
Interfaces to Gemini System |
11.5 |
Inspection |
… |
|
REQ-FPR-0506 |
Gemini Supplied ICDs |
11.6 |
… |
… |
|
REQ-FPR-0507 |
Gemini Supplied CICS |
11.7 |
… |
… |
|
REQ-FPR-0508 |
Gemini Supplied OIWFS I/F |
11.8 |
… |
… |
|
REQ-FPR-0509 |
Gemini Supplied DHS |
11.9 |
… |
… |
|
REQ-FPR-0510 |
Engineering Interface |
11.10 |
Demonstration |
… |
|
REQ-FPR-0511 |
Engineering I/F Physical I/F |
11.11 |
Inspection |
… |
|
REQ-FPR-0512 |
Engineering Interface User I/F |
11.12 |
Demonstration |
… |
|
REQ-FPR-0513 |
Eng I/F Command and Control |
11.13 |
Laboratory test |
Base Facility test |
|
REQ-FPR-0514 |
Engineering I/F Data Capture |
11.14 |
Laboratory test |
Base Facility test |
|
REQ-FPR-0600 |
ISS Interface |
12.1 |
Inspection |
Demonstration |
|
REQ-FPR-0601 |
ISS Ports |
12.2 |
Inspection |
Demonstration |
|
REQ-FPR-0602 |
Mounting Repeatability |
12.3 |
… |
Telescope test |
|
REQ-FPR-0603 |
Mounting Plate Material |
12.4 |
Inspection |
… |
|
REQ-FPR-0604 |
Fasteners |
12.5 | ||